Clinical Review

Biocompatibility: a biomechanical and biological concept in total hip replacement

I.D. Learmonth
University Department of Orthopaedic Surgery, Level 5, Bristol Royal Infirmary, Bristol, BS2 8HW, UK 

Correspondence to: I.D. Learmonth, University Department of Orthopaedic Surgery, Level 5, Bristol Royal Infirmary, Bristol, BS2 8HW, U.K. Email: ian.learmonth@bristol.ac.uk

              

Introduction

Biomechanical compatibility

Biological response

 

 

Discussion

References

 

 

Keywords: Biocompatibility, biomechanical, biological, hip, particle debris
Surg J R Coll Surg Edinb Irel., 1 February 2003, 1-8

The insertion of any implant or prosthesis into bone usually changes the biomechanical environment and, thus, alters the stresses and strains applied to the bone. Both bone overload and excessive stress protection can result in bone resorption. The material and geometry of any implant should be designed to avoid excessive flexural mismatch. Incompatibility of materials may result in interface and mechanical failure, with the consequent generation of particulate debris, subsequent osteolysis and implant failure. Particulate debris can be generated from articulating surfaces and from any other modular or fixation interface. Larger particles are associated with a foreign body giant cell reaction. Polyethylene particles in the size range of 0.5 to 10µm excite a cytochemical reaction that culminates in osteolysis. The precise pathogenesis of osteolysis has not been characterised, but it is probable that different pathogenetic mechanisms are involved in the different radiological types of osteolysis. A large number of very small metallic particles are released from metal-on-metal couples. These may cause mutagenic damage (chromatid breaks, chromosome translocations, aneuploidy, etc.). In defining implant biocompatibility it is essential to consider the biological response both to an altered mechanical environment and to the liberation of particulate debris

INTRODUCTION
Orthopaedic implants inserted in the 19th century and the first half of the 20th century were destined to failure, both as a result of metallic corrosion and of structural failure. Advances in material sciences and bio-engineering greatly improved the quality of orthopaedic implants, and structural failure of the implants became the exception rather than the rule. However, having reduced the incidence of failure, it became necessary to consider the biological implications of metal or plastic prostheses that survive in the longer term. 

Biomaterials should be non-toxic, both locally and systemically, should not be allergenic or carcinogenic and should be bio-acceptable - in particulate and bulk form as well as the breakdown products. In addition, the implant should ideally mimic the tissue it is replacing as closely as possible.

Inserting a metal prosthesis down the medullary cavity of a femur alters the loading pattern and, thus, the stresses and strains in the proximal femur. In accordance with Wolff’s law, this may result in bone remodelling and reduction in bone density.1 Hence, it is important to evaluate the effects of the altered biomechanics in bone caused by the insertion of a stiff metal prosthesis. It is also important to recognise that different materials have different properties - modulus of elasticity, scratch sensitivity, hardness, etc. Certain materials may be incompatible when intermittently loaded in juxtaposition in vivo.

While most materials used today are biocompatible in bulk, they may excite a variable response when presented to the tissues in particulate form. Wear at the articular interface of a metal-plastic couple accounts for the generation of a large number of polyethylene particles (millions of particles may be liberated with each step). The smaller of these particles excite a cytochemical response. Metal particles may be released as a result of corrosion and abrasion. Large numbers of much smaller metal particles are produced which may cause genetic damage.

This article reviews the implications of these considerations in more detail.

Figure 1: (a [above] and b [below]): A finite element analysis model shows the reduced stress protection in the calcar region as the porous coating is reduced from one ninth (a) to one third (b)

BIOMECHANICAL COMPATIBILITY

Implant design
There was a relatively high incidence of fracture of the femoral stems of early Charnley Low Friction Arthroplasties.2 However, while changing the cross-sectional geometry and increasing the dimensions reduced the incidence of fracture, it resulted in a stiffer stem and the resultant flexural mismatch with the surrounding cement and bone was associated with an increased incidence of aseptic loosening.3 It should be remembered that the flexibility of an implant depends both on the material composition and the cross-sectional geometry.4 If the stem of a femoral component makes contact with the diaphyseal cortex there is potential for distal load transfer and proximal stress shielding with resultant stress protection bone resorption. In these circumstances, flexibility of the implant has been inversely related to thigh pain and stress shielding.5,6 Load transfer is enhanced if the stem has surface treatment which encourages bone ongrowth (osseointegration).7 The more distal the load is transferred, the greater the potential for proximal stress protection.8 To investigate this a two-dimensional model of a cementless stern inserted into the proximal femur was created using finite elememt analysis. As the extent of the metaphyseal porous coating was reduced proximally, a commensurate reduction in the extent of reduced strain was noted in the calcar region. This exactly mirrored the demineralisation in the calcar region encountered in vivo when a stiff proximally coated cementless stem was used at total hip replacement (Figure 1 a-d). A fully coated stiff cementless stem may result in stress protection of sufficient severity to significantly compromise proximal femoral bone stock and complicate any subsequent revision surgery.

Figure 1: (c [above] and d below]): This reflects the clinical situation. (c) Early post-operative radiograph. (d) The rounding off and "cancellisation" of the calcar that takes place after two years

Various strategies have been used in an attempt to reduce the stiffness of the stem, including slots and the use of composite materials. However, the linear evaluation of bone density in the post-operative period using DEXA has not identied any real difference in the change in periprosthetic bone density with contemporary cementless implants, when compared with earlier generation implants.9

Cortical hypertrophy is a further example of an osseous response to altered mechanical environment. In a recent survey of 109 cemented Muller stems, 43% were found to present with distal cortical hypertrophy. The precise significance of this remains unknown, and there was no correlation with thigh pain.10,11 Specific design features of an implant may predispose to early aseptic loosening and failure. For example, the rounded medial and lateral surfaces below the shoulder of the flanged 3M Capital hip are thought to have predisposed to reduced torsional stability and an increased incidence of aseptic loosening.12 Conversely, however, the long-term survival of an implant may depend on the pattern of loading and the associated bone response.13,14

Materials
Care should also be taken to ensure the compatibility of different materials. This relates not only to potential electrochemical interaction, but also to their physical compatibility. Jonck and Grobbelaar (1990, 1992) investigated the biocompatibility of a bioglass-glass ionomer cement in baboons.15,16 The material was found to be very biocompatible. However, when used as a bone-cement at total hip replacement, the differential in hardness between the ceramic and the relatively soft titanium stem resulted in catastrophic wear of the stem (Figure 2).

Figure 2: Severe abrasive and erosive damage to the femoral stem

Impingement or abrasive movement between any two materials of very different hardness is always associated with the risk of very severe wear.17 The femoral neck of the hip prosthesis shown in Figure 3 was impinging against the ceramic cup, and was found to be heavily notched superiorly. This would have released a considerable amount of metallic debris into the tissues and if this implant had not been revised for aseptic loosening, the neck would almost certainly have fractured with the fullness of time.

Figure 3: The deep notch on the femoral neck (caused by impingement against a ceramic socket)

Simulators are extensively used to test the strength and durability of joint prostheses. While it is clearly critical to avoid structural failure of the implant, it is equally important to consider the interaction between the prosthesis and other materials in the composite reconstruction. In addition, cognisance must be taken of the biological response to the altered biomechanical environment. Furthermore, simulators should replicate the loading profile of activities of daily living, and should produce wear debris with a morphology of similar type and amount to that produced in vivo.18  It is then necessary to consider the biological response to this particulate debris.

BIOLOGICAL RESPONSE TO PARTICULATE DEBRIS OSTEOLYSIS

Osteolysis
In 1987, Jones and Hungerford coined the term “cement disease” to depict the extensive osteolysis encountered with aseptic loosening of cemented total hip replacements.19 However, the aggressive osteolysis seen after cementless total hip replacement proved this to be a misnomer and the term particulate disease was coined.

It is now widely accepted that ultrahigh molecular weight polyethylene is probably the major culprit contributing to particle-mediated osteolysis.20 The majority of these particles are submicron in size and are engulfed by cells which are then stimulated to produce mediators of bone resorption.21,22 The pseudosynovial membrane (PSM) encountered at the component-bone interface of loose prosthesis contains a number of these cells, including macrophages, fibroblasts and lymphocytes. 23,24

Conditioned media (CM) removed from cultures of PSM have been shown to induce bone release in vitro.23,25 Perry et al (1996) demonstrated a relationship between the bone resorbing ability of CM obtained from cultures of periprosthetic tissues and their levels of bone remodelling agents. Bone resorbing activity was determined by measuring the release of Ca++ from45 pre-labelled mouse calvaria in organ culture, while the cytokine prostanoid levels in the CM were measured by immuno-assay (interleukin [IL]-1ß, IL-6, tumour necrosis factor a [TNFa] and prostaglandin [PG]E2). In general, significantly higher levels of cytokines were encountered in those CM with bone resorbing activity. However, interestingly, neither dialysis of the CM nor the addition of neutralising sera to IL-1ß and IL-1a, either alone or in combination, reduced the bone resorbing activity of the CM. This led Perry et al (1996) to conclude that while these agents may act synergistically to mediate osteolysis around failed joint implants, other unidentifed bone resorbing agents must be involved.26

Santavirta et al (1990) identifed an aggressive form of osteolysis, which they ascribed to an unlinking of the normal inflammatory response.27 Schmalzried et al (1992) have described different types of osteolysis - linear, lytic and mixed.28 I believe two further types can be added - discrete osteolysis and “pseudotumour”. The different descriptive radiological types of osteolysis are shown in Figure 4. It is difficult to believe that they all have the same aetiopathogenic mechanism.

Perry et al (1997) measured the levels of bone remodelling agents in CM from cultures of PSM and correlated these with the radiographic appearances of the failed joint implants.29 Significant correlations of specific bone remodelling agents were found in different types of osteolysis. It was suggested, therefore, that co-regulation of these bone remodelling agents differs with the radiographic appearances of the failed joint implants.

Other factors have been incriminated in osteolysis. These include leukaemia inhibitory factor (LIF), oncostatin M, atypical fibroblasts (similar to those seen in the pannus of rheumatoid arthritis) and increased joint pressure. Although the precise aetiopathogenesis of osteolysis remains unclear, it is evident that particulate debris plays a central role. Abnormal stresses and strains associated with loosening and abnormal movement may be an aggravating factor, but are clearly not responsible for initiating the process. It is necessary, therefore, to explore means of reducing the particulate load and preventing access to the fixation interface. It may also be possible to modulate the biological response pharmacologically (e.g. with bisphosphonates).30

 

Figure 4a

               

                                                        Figure 4b                                                        Figure 4c                                            Figure 4d

Figure 4: Different radiological types of osteolysis: (a) discrete (b) linear (c) erosive (d) mixed 

 

 

 

                              

                             Figure 4ei                                           Figure 4eii                                             Figure 4eiii                                    Figure 4iv

Figure 4: Different radiological types of osteolysis: (e) i-iv pseudotumour: (i) Early post-operative radiograph. (ii) Five-year follow-up. (iii-iv) The appearances shown on the AP and lateral radiographs at eight years are reminiscent of a malignant neoplasm

Metallic debris
Modular metal interfaces and corrosion and abrasion, as a result of differential micromovement, liberate a large amount of metallic debris into the system. While the volumetric wear of a metal-on-metal articulation is much lower than that of metal-on-plastic, it produces a far greater number of particles.31 These particles are very much smaller than the size of the average plastic particle, and do not excite an inflammatory response. However, metallic particles do accumulate in cells and may cause mutagenic damage. Metal particles have been identifed in tissues adjacent to the implant, in the regional and distant lymph nodes, liver and spleen.32,33  In addition, increased levels of trace metals have been reported in the serum and urine especially if there is loosening of the prosthesis.34,35

Hypersensitivity has been well documented following total hip replacement, particularly with stainless steel implants. The prevalence of metal sensitivity in eight studies varied between 3% and 43% (average +/- 25%) - approximately double the prevalence found in the normal population. The prevalence documented in seven publications reporting metal sensitivity in poorly functioning implants varied between 13% and 71% (average 60%). This was five times greater than that found in the general population.

Using inductively coupled mass spectrometry, Case et al (1994) measured metal levels in the body (synovium, pelvic lymph nodes, para-aortic lymph nodes, liver, spleen etc) at post-mortem and compared them with controls.36  The lymph node concentration was many times higher than in the control cases - as much as cobalt X 2000, chromium X 362, molybdenum X 58 and iron X 15. The material which is disseminated is not inert and may provoke cytokine-mediated bone resorption and nuclear pyknosis and cell death.37-39 Necrosis and fibrosis of local lymph nodes could compromise the local immune response and impair the host response to infection. Indeed, Bravo et al (1990) have demonstrated inhibition of T cell activities in tissue culture.40

Case et al (1996) analysed blood and bonemarrow samples from 71 patients at revision arthroplasty of a loose or worn prosthesis, and 30 control patients at primary arthroplasty with cytogenetic and molecular biological techniques. 41 There was a higher chromosomal aberration rate in cells adjacent to the prosthesis at revision surgery, compared with iliac crest biopsies from the same patients or with femoral bone marrow at primary arthroplasty.

Recognising that the long-term biological effects of wear debris were largely unknown, Doherty et al (2001) investigated whether there was any evidence of cumulative mutagenic damage in peripheral blood lymphocytes of patients undergoing revision arthroplasty of predominantly metal-on-metal hip replacements, compared with those at primary arthroplasty (Figure 5).42 Using a chromosome painting technique they demonstrated a threefold increase in aneuploidy and a two-fold increase in chromosomal translocations that could not be explained by the confounding variables of smoking, gender, age and diagnostic radiographs. In patients with Ti Va A1 prostheses they noted a five-fold increase in aneuploidy but no increase in chromosomal translocations. By contrast, in patients with cobalt chrome prostheses there was a 2.5-fold increase in aneuploidy and a 3.5-fold increase in chromosomal translocations. In six patients with stainless steel prostheses there was no increase in either aneuploidy or chromosomal translocations.

Figure 5a

Figure 5b

Figure 5: (a) A reciprocal translocation between chromosomes 1 and 2 (b) There are three copies of the chromosome I to gove 47 chromosomes. This is aneuploidy

It is clear, therefore, that metallic debris does provoke a biological response, and the nature of the response is influenced by the type of alloy used in the prosthesis. These factors need to be considered when assessing the long-term implications of joint replacement.

DISCUSSION
In considering a new prosthetic design, strategies should be formulated to ensure that biocompatibility embraces both biomechanical and biological imperatives. We are currently evaluating a metaphyseal loading implant to assess how it affects the pattern of loading in the proximal femur, quantitatively and qualitatively. More conservative options (such as resurfacing arthroplasty) are likely to provide more normalised loading regimens in the proximal femur.

It has been shown that a highly polished double tapered stem converts the shear stresses (in the cement, in the bone and at the fixation interface) into compressive stresses. This beneficial effect is sustained by the ability of the prosthesis to subside and re-engage the taper - thus restoring the compressive loading regimen. It is probable that the beneficial inuence of the stem design on the mechanical environment has contributed to the excellent 10-year results reported recently.43

Wroblewski et al (2001) have recently observed that reduced loading of the calcar is associated with loss of bone stock in the long-term using the Charnley Low Friction Arthroplasty. They suggested that using a cemented stem with a triple taper will provide more effective loading of the calcar in the longer term - and their seven-year results are most encouraging.44

It would be highly desirable to have a model that was predictive of the biological effects of altering the biomechanical environment. This could perhaps be achieved by linear monitoring of the movement (RSA) and bone density changes (DEXA) around an implant and then iteratively and dynamically evolving a finite element analysis model. This could provide validation to a predictive theoretical model that is often .awed by untenable assumptions. Aseptic loosening is the commonest cause of failure of total joint replacement, and osteolysis is the commonest cause of aseptic loosening. It behoves us, therefore, to address this issue. Current strategies to reduce the particulate load include minimising the generation of debris by optimising the materials, design and surgical implantation of the prosthesis. In addition, an attempt should be made to seal the interfaces (cement, circumferential bioactive coatings etc.) and so prevent particle access and ingress. A better understanding of the aetiopathogenesis of osteolysis would allow us to pharmacologically manipulate and modulate the biological response to the debris.

The success of total joint replacement has created its own nemesis. Indications have been relaxed, and joint replacement is increasingly being carried out in younger cohorts of patients. Particle-driven disease (asbestosis, etc.) only becomes clinically apparent after a long-term interval. The adverse experience with silicone in North America (although somewhat nebulous) suggests that there is no room for complacency. There can be little doubt about the carcinogenic potential of metallic substances, particularly chromium, cobalt and nickel, released from implants.45 However, it should be stressed that the fact that there are mutagenic changes in peripheral blood lymphocytes does not prove that there will be clinical expression of disease in the patient in the long-term. Of perhaps greater significance is the emerging evidence that organometallic ions produced by implants constitute a chronic stress on the immune system.46 Donati et al (1998) have demonstrated a selective decrease in the numbers of T and B lymphocytes and natural killer cells in response to increased levels of chromium, cobalt and nickel ions.47  The association between implant-generated wear debris-and the development of disease remains uncertain and controversial.

It is clearly essential to establish ongoing communication between clinicians, scientists and manufacturers. A surgeon should be as knowledgeable about the biological implications of the material and prosthesis he/she is implanting as the physician is about the side effects of any drug he prescribes.

REFERENCES
1. Wolff I. Das Gosetz derTransformation der Knochen. Berlin, Hirschwald, 1982
2. Charnley J. Fracture of femoral prostheses in total hip replacement. A critical study. Clin Orthop 1975; 111: 105-20
3. Dall DM, Learmonth ID, Solomon MI, Miles AW, Davenport JM. Fracture and loosening of Charnley femoral stems. Comparison between first generation and subsequent designs. J Bone Joint Surg 1993; 75-B: 259-65
4. Sarmiento A and Gruen TA. Radiographic analysis of a low-modulus titanuium alloy total hip component. Two to six-year follow-up. J Bone Joint Surg 1985; 67-A: 48-56
5. Franks E, Mont MA, Maar DC, Jones LC, Hungerford DS.Thigh pain as related to bending rigidity of the femoral prosthesis and box. Trans Orthop Res Soc 1992; 17: 296
6. Bobyn ID, Glassman AH, Goto H, Krygier JJ, Miller IE, Brookes CE. The effect of stem stiffness on femoral bone resorption after canine porous coated hip arthroplasty. Trans Orthop Res Soc 1991; 16: 502
7. Engh CA, Bobyn ID, Glassman AH. Porous-coated hip replacement, the factors governing bone ingrowth, stress shielding and clinical results. J Bone Joint Surg 1987; 69-B: 45-55
8. Engh CA, McGovern TF, Engh CA Jr, Macalino GE. Clinical experience with the Anatomic Medullary Locking (AML) prosthesis for primary total hip replacement. BF Morrey (Ed.). Biological, material, and mechanical considerations of joint replacement. New York: Raven Press, 1993: 167-84
9. Eastaugh-Waring SJ, Eldridge JE, Lee M, Learmonth ID.Remodelling around three differing designs of cementless femoral stems. Hip International 2002; 12: 197
10. Katsimihas M, Islam R, Lee MB, Learmonth ID.Development of distal cortical hypertrophy and predisposing factors following total hip replacement. The medium to long-term radiographic assessment of the 32mm Muller femoral prosthesis. Hip International 2002; 12: 196-97
11. Pearse MF, Karachalios T, Langkamer VG, Atkins RM, Solomon L. Femoral remodelling following total hip replacement. J Orth Rheum 1992; 5: 85-96
12. 3MTM Capital Hip System: the lessons learned from an investigation. Royal College of Surgeons of England,2001
13. Oh I, Harris WH. Proximal strain distribution in the loaded femur. J Bone Joint Surg 1978; 60-A: 75-85
14. Ling RSM. Observations on the fixation of implants to the bony skeleton. Clin Orthop 1986; 210: 80-96
15. Jonck LM, Grobbelaar CJ. Ionos bone cement ( glassionomer): an experimental and clinical evaluation in joint replacement. Clin Mater 1990; 6: 323-59
16. Jonck LM, Grobbelaar CJ. A glass-ionomer for reconstructive surgery. Ionogran - an ionomeric micro implant. A biological evaluation. Clin Mater 1992; 9: 85-103
17. Von Recum A (Ed.). Handbook of biomaterials evaluation. New York, Macmillan, 1986
18. Santavirta S, Konttinen YT, Lappalainen R, Anttila A, Goodman SB, Lind M, Smith L, Tagaki M, Gomez-Barrena E, Nordsletten L, Xu I-W. Materials in total joint replacement.Curr Orthop 1998; 12: 51-57
19. Jones LC, Hungerford DS. Cement disease. Clin Orthop 1987; 225: 192-206
20. Friedman IR, Black J, Galante JO, Jacobs JJ, Skinner HB.Current concepts in orthopaedic biomaterials and implant fixation. J Bone Joint Surg 1993; 75-A: 1086-1109
21. Margevicius KJ, Bauer TW, McMahon IT, Brown SA Merritt K. Isolation and characterisation of debris from around total joint prostheses. J Bone Joint Surg 1994; 76-A: 1664-75
22. Shanbhag AS, Jacobs JJ, Glant TT, Gilbert JL, Black J, Galante JO. Composition and morphology of wear debris in failed uncemented total hip replacement. J Bone Joint Surg 1994; 76-B: 60-7
23. Goldring SR, Jasty M, Roelke MS, Rourke CM, Bringhurst FR, Harris WH. Formation of a synovial-like membrane at the bone cement interface. Arthritis Rheum 1986; 29:836-42
24. Perry MJ, Mortuza FY, Ponsford FM, Elson CJ, Atkins RM. Analysis of cell types and mediator production from tissues surrounding loosening joint implants. Br J Rheumatol 1995; 34: 1127-34
25. Appel AM, Sowden WG, Siverhus SW, Hopson CN, Herman JH. Prosthesis-associated psuedomembrane induced bone resorption. Br J Rheumatol 1990; 29: 32-6
26. Perry MJ, Ponsford FM, Mortuza FY, Learmonth ID, Atkins RM and Elson CJ. Osteolytic properties of the synovial-like tissue from aseptically failed joint prostheses. Br J Rheumatol 1996; 35: 943-50
27. Santavirta S, Hoikka V, Eskola A, Konttinen YT, Paavilainen T, Tallroth K. Aggressive granulomatous lesions in cementless total hip arthroplasty. J Bone Joint Surg 1990; 72-B: 980-84
28. Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty. J Bone Joint Surg 1992; 74-A: 849-63
29. Perry MJ, Mortuza FY, Ponsford FM, Elson CJ, Atkins RM, Learmonth ID. Properties of tissue from around cemented joint implants with erosive and/or linear osteolysis. J Arthroplasty 1997; 12: 670-76
30. Goodship AE, Lawes TJ, Green J, Eldridge, JD. Bisphosphonates can inhibit mechanically related loosening of hip prostheses. J Bone Joint Surg 1999; 81-B, Supp III: 319
31. Firkins PJ, Tipper JL, Saadetzadeh MR, Ingham E, Stone MH, Farrar R, Fisher J. Quantitative analysis of wear and wear debris from metal on metal hip prostheses tested in a physiological hip joint simulator. Biomed Mater Eng 2001; 11: 143-57
32. Revell PA. Tissue reactions to joint prostheses and the products of wear and corrosion. In Berry CL (Ed.) Bone and Joint Disease. Berlin: Springer, 1982:73-101
33. Gray MH, Talbert ML, Talbert WM, Bansal M, Hsu A. Changes seen in lymph nodes draining the sites of large joint prostheses. Am J Surg Pathol 1989; 13:1050-56
34. Coleman RF, Herrington J, Scales IT. Concentration of wear products in hair, blood and urine after total hip replacement. Br Med J 1973; 1:527-29
35. Sunderman FW Jr, Hopfer SM, Swift T et al. Cobalt, chromium and nickel concentrations in body fluids of patients with porous-coated knee or hip prostheses. J Orthop Res 1989; 7: 307-15
36. Case CP, Langkamer WG, James C, Palmer MR, Kemp AJ, Heap PF, Solomon L. Widespread dissemination of metallic wear debris from implants. J Bone Joint Surg 1994; 76-B:706-12
37. Spector M, Shortkroff S, Hsu H-P, Lane N, Sledge CB, Thornhill TS. Tissue changes around loose prostheses: a canine model to investigate the effects of an anti-inflammatory agent. Clin Orthop 1990; 261:140-52
38. Galante JO, Lemons J, Spector M, Wilson PD Jr, Wright TM. The biologic effects of implant materials. J Orthop Res 1991;9:760-75
39. Howie DW. Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty 1990; 5:337-48
40. Bravo I, Carvaltro GS, Barbosa MA, de Sousa M. Differential effects of eight metal ions on lymphocyte differentiation antigens in vitro. J Biomed Mater Res 1990; 24:1059-68
41. Case CP, Langkamer VG, Howell RT, Webb J, Standen G, Palmer M, Learmonth ID. Preliminary observations on possible premalignant changes in bone marrow adjacent to worn total hip arthroplasty implants. Clin Orthop 1996; 329S, S269-79
42. Doherty AT, Howell RT, Ellis LA, Bisbinas I, Learmonth ID, Newson R, Case CP. Increased chromosome translocations and aneuploidy in peripheral blood lymphocytes of patients having revision arthroplasty of the hip. J Bone Joint Surg 2001; 83-B: 1075-81
43. Williams HDW, Browne G, Gie GA, Ling RSM, Timperley AJ, Wendover NA. The Exeter universal cemented femoral component at 8 to 12 years. J Bone Joint Surg 2002; 84-B:324-34
44. Wroblewski BM, Siney PD, Fleming PA. Triple taper polished cemented stem in total hip arthroplasty: rationale for the design, surgical technique, and 7 years of clinical experience. J Arthroplasty 2001; 16(8 Supp. 1): 37-41
45. Sunderman FW. Mechanisms of metal carcinogenesis. Biol Trace Elements Res 1979; 1: 63-86
46. Wang JY, Tsukayama DT, Wicklund BH, Gustilo RB. Inhibitors of T and B cell proliferation by titanium, cobalt and chromium: role of IL-1 and IL-6. J Biomed Mater Res 1996; 32:655-61
47. Donati ME, Savarino L, Granchi D, Ciapetti G, Cervellati M, Rotino R, Pizzoferrato A. The effects of metal corrosion debris on immune system cells. Chir Organi Mov 1998; 83:387-93

Copyright: 12 August 2002